http://www.cnr.it/ontology/cnr/individuo/prodotto/ID48357
Dependence of calibration sensitivity of a polysulfone/Ru(II)-Tris(4,7-diphenyl-1,10-phenanthroline)-based oxygen optical sensor on its structural parameters (Articolo in rivista)
- Type
- Label
- Dependence of calibration sensitivity of a polysulfone/Ru(II)-Tris(4,7-diphenyl-1,10-phenanthroline)-based oxygen optical sensor on its structural parameters (Articolo in rivista) (literal)
- Anno
- 2008-01-01T00:00:00+01:00 (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#doi
- 10.1016/j.aca.2008.08.019 (literal)
- Alternative label
Badocco, D.; Mondin, A.; Pastore, P.; Voltolina, S.; Gross, S. (2008)
Dependence of calibration sensitivity of a polysulfone/Ru(II)-Tris(4,7-diphenyl-1,10-phenanthroline)-based oxygen optical sensor on its structural parameters
in Analytica chimica acta (Print)
(literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#autori
- Badocco, D.; Mondin, A.; Pastore, P.; Voltolina, S.; Gross, S. (literal)
- Pagina inizio
- Pagina fine
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroVolume
- Rivista
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#pagineTotali
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroFascicolo
- Note
- ISI Web of Science (WOS) (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#affiliazioni
- [Pastore, Paolo; Gross, Silvia] Univ Padua, Dept Chem Sci, ISTM CNR, I-35131 Padua, Italy.
[Gross, Silvia] INSTM UdR Padua, I-35131 Padua, Italy. (literal)
- Titolo
- Dependence of calibration sensitivity of a polysulfone/Ru(II)-Tris(4,7-diphenyl-1,10-phenanthroline)-based oxygen optical sensor on its structural parameters (literal)
- Abstract
- The optimum performance of an optical oxygen sensor based on
polysulfone (PSF)/[Ru(II)-Tris(4,7-diphenyl-1,10-phenanthroline)]
octylsulfonate (Ru(dpp)OS) was checked by carefully tuning the
parameters affecting the membrane preparation. In particular, membranes
having thickness ranging between 0.2 and 8.0 mu m with various
luminophore concentrations were prepared by dip-coating and tested. The
membrane thickness was controlled by tuning the solution viscosity, and
was measured both by secondary ion mass spectrometry (SIMS) and by
visible spectroscopy (Vis). Luminescence-quenching-based calibration
was a single value of the Stern-Volmer constant (K-SV') for membranes
containing up to 20 mmol Ru(dpp) g(-1) PSF (1.35 mu m average
thickness). The K-SV' value decreased for larger-concentration. The
highest sensitivity was obtained with membrane thickness around 1.6 mu
m, having a response time close to 1 s. Thicker membranes exhibited an
emission saturation effect and were characterized by longer response
time. The K-SV' behavior was interpreted on the basis of a mathematical
approach accounting for the contribution of luminescence life-time
(tau(0)), oxygen diffusion coefficient (D-O2) and oxygen solubility
inside the membrane (S-O2) establishing the role of all of them and
allowing their experimental determination. Moreover, a simple
experimental way to estimate K-SV' without needing calibration was
proposed. It was based either on the light emission asymmetry or on the
percent variation of light emission on passing from pure nitrogen to
pure oxygen. (C) 2008 Elsevier B.V. All rights reserved. (literal)