http://www.cnr.it/ontology/cnr/individuo/prodotto/ID30210
Measurement of electrode current pulses from cochlear implants (Articolo in rivista)
- Type
- Label
- Measurement of electrode current pulses from cochlear implants (Articolo in rivista) (literal)
- Anno
- 2005-01-01T00:00:00+01:00 (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#doi
- 10.1109/TIM.2005.855120 (literal)
- Alternative label
Tognola G., Burdo S., Caponio M., Norgia M., Parazzini M., Ravazzani P., Grandori F., Svelto C. (2005)
Measurement of electrode current pulses from cochlear implants
in IEEE transactions on instrumentation and measurement
(literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#autori
- Tognola G., Burdo S., Caponio M., Norgia M., Parazzini M., Ravazzani P., Grandori F., Svelto C. (literal)
- Pagina inizio
- Pagina fine
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#url
- http://ieeexplore.ieee.org/xpl/articleDetails.jsp?arnumber=1514670 (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroVolume
- Rivista
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroFascicolo
- Note
- copu (literal)
- ISI Web of Science (WOS) (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#affiliazioni
- Tognola G.: Istituto di Ingegneria Biomedica ISIB CNR
Burdo S.: Servizio di Audiovestibologia, Ospedale di Circolo di Varese, 21100 Varese, Italy
Caponio M.: Istituto di Ingegneria Biomedica ISIB CNR e Servizio di Audiovestibologia, Ospedale di Circolo di Varese, 21100 Varese, Italy
Norgia M.: Dipartimento di Elettronica e Informazione, Politecnico di Milano
Parazzini M.: Istituto di Ingegneria Biomedica ISIB CNR
Ravazzani P.: Istituto di Ingegneria Biomedica ISIB CNR
Grandori F.: Istituto di Ingegneria Biomedica ISIB CNR
Svelto C.: Dipartimento di Elettronica e Informazione, Politecnico di Milano (literal)
- Titolo
- Measurement of electrode current pulses from cochlear implants (literal)
- Abstract
- The experimental setup for the measurement of the current pulses generated at the different electrodes of a cochlear implant is presented. The electrical stimuli were modulated by an ad hoe software, transferred through a custom-made cochlear-implant programming interface to the speech processor, encoded as radio frequency (RF) signals, delivered via coil to the implant receiver/stimulator, and hence sent as current pulses to the electrode array of the implant. In the measurement procedure, each electrode of the implant array is connected to an external port, and the current waveform at any single electrode can be measured by a probe and a digital oscilloscope. It was shown that the Nucleus Cochlear Implant delivers current pulses according to specifications with emulated clinical load resistances within the physiological range. (literal)
- Prodotto di
- Autore CNR
- Insieme di parole chiave
Incoming links:
- Prodotto
- Autore CNR di
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#rivistaDi
- Insieme di parole chiave di