Integrated SQUID sensors for low cross-talk multichannel systems (Articolo in rivista)

Type
Label
  • Integrated SQUID sensors for low cross-talk multichannel systems (Articolo in rivista) (literal)
Anno
  • 2006-01-01T00:00:00+01:00 (literal)
Alternative label
  • C Granata, A Vettoliere, M Luiso and M Russo (2006)
    Integrated SQUID sensors for low cross-talk multichannel systems
    in Journal of physics. Conference series (Print)
    (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#autori
  • C Granata, A Vettoliere, M Luiso and M Russo (literal)
Pagina inizio
  • 1235 (literal)
Pagina fine
  • 1238 (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroVolume
  • 43 (literal)
Rivista
Note
  • ISI Web of Science (WOS) (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#affiliazioni
  • Istituto di cibernetica \"Edoardo Caianiello\" (literal)
Titolo
  • Integrated SQUID sensors for low cross-talk multichannel systems (literal)
Abstract
  • We present a fully integrated dc-SQUID magnetometer based on niobium technology including a new feedback coil design. In respect to a standard SQUID design, such a feedback-coil design was optimized in order to reduce the mutual inductance with the neighbours and to increase the coupling with the pick-up coil of the SQUID itself. In such a way, it is possible to reduce cross-talks due to both feedback coil and wires. Experimental results about the characterization of the device and the crosstalk measurements are reported. The measurements have been performed in liquid helium using a low noise readout electronics specifically designed for large multichannel SQUID based instrumentations. The experimental data have shown a substantial reduction of cross-talk between neighbouring sensors. (literal)