Modelling in the study of interaction of Hemopump device and artificial ventilation (Articolo in rivista)

Type
Label
  • Modelling in the study of interaction of Hemopump device and artificial ventilation (Articolo in rivista) (literal)
Anno
  • 2006-01-01T00:00:00+01:00 (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#doi
  • 10.1016/j.compbiomed.2005.08.001 (literal)
Alternative label
  • De Lazzari C.; Darowski M.; Ferrari G.; Pisanelli D.; Tosti G. (2006)
    Modelling in the study of interaction of Hemopump device and artificial ventilation
    in Computers in biology and medicine; Pergamon-Elsevier Science Ltd., Oxford (Regno Unito)
    (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#autori
  • De Lazzari C.; Darowski M.; Ferrari G.; Pisanelli D.; Tosti G. (literal)
Pagina inizio
  • 1235 (literal)
Pagina fine
  • 1251 (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroVolume
  • 36 (literal)
Rivista
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#note
  • In: Computers in Biology and Medicine, vol. 36 (11) pp. 1235-1251. Elsevier, 2006. (literal)
Note
  • PubMe (literal)
  • ISI Web of Science (WOS) (literal)
  • Scopu (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#affiliazioni
  • C.N.R., Institute of Clinical Physiology, Cardiovascular Engineering Department, Rome, Italy P.A.N. Institute of Biocybernetics and Biomedical Engineering, Center of Excellence ARTOG, Warsaw, Poland C.N.R., Institute for Cognitive Science and Technology, Rome, Italy (literal)
Titolo
  • Modelling in the study of interaction of Hemopump device and artificial ventilation (literal)
Abstract
  • The aim of this work is to evaluate in different ventricular conditions the influence of joint mechanical ventilation (MV) and Hemopump assistance. To perform this study, we used a computer simulator of human cardiovascular system where the influence of MV was introduced changing thoracic pressure to positive values. The simulation confirmed that haemodynamic variables are highly sensitive to thoracic pressure changes. On the other hand, Hemopump assistance raises, among the others, mean aortic pressure, total cardiac output (left ventricular output flow plus Hemopump flow) and coronary flow. The simulation showed that the joint action of Hemopump and positive thoracic pressure diminishes these effects. (literal)
Editore
Prodotto di
Autore CNR
Insieme di parole chiave

Incoming links:


Autore CNR di
Prodotto
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#rivistaDi
Editore di
Insieme di parole chiave di
data.CNR.it