Towards new temperature standards for contact thermometry above 6600C (Articolo in rivista)

Type
Label
  • Towards new temperature standards for contact thermometry above 6600C (Articolo in rivista) (literal)
Anno
  • 2002-01-01T00:00:00+01:00 (literal)
Alternative label
  • Marcarino P., Merlone A. (2002)
    Towards new temperature standards for contact thermometry above 6600C
    in Metrologia (Print)
    (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#autori
  • Marcarino P., Merlone A. (literal)
Pagina inizio
  • 395 (literal)
Pagina fine
  • 398 (literal)
Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroVolume
  • 39 (literal)
Rivista
Note
  • ISI Web of Science (WOS) (literal)
Titolo
  • Towards new temperature standards for contact thermometry above 6600C (literal)
Abstract
  • In the course of a study on the behaviour of coupled heat-pipes, perfect agreement has been obtained for the first time between measurements on the immersion characteristics in a mercury heat-pipe and the Clausius- Clapeyron profile. This result opens the perspective of using coupled heat- pipes to relate thermodynamically two different temperature ranges, with the possibility of redefining one temperature range in terms of another. As a consequence, using sodium for the second heat-pipe, the very high reproducibility of standard platinum resistance thermometers(SPRTs) below 400 °C can be transferred to the temperature range above 660 °C, surpassing the characteristics of high-temperature platinum resistance thermometers (HTSPRTs). (literal)