http://www.cnr.it/ontology/cnr/individuo/prodotto/ID32106
Structural polymer composite materials for muscle-skeletal diagnostic systems (Articolo in rivista)
- Type
- Label
- Structural polymer composite materials for muscle-skeletal diagnostic systems (Articolo in rivista) (literal)
- Anno
- 2010-01-01T00:00:00+01:00 (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#doi
- 10.1177/0731684410391514 (literal)
- Alternative label
M.F. Pernice1, D. Annicchiarico2, A. Petriccione2,
G. De Tommaso1, M. Meo1,3, V. Antonucci2 and M. Zarrelli2 (2010)
Structural polymer composite materials for muscle-skeletal diagnostic systems
in Journal of reinforced plastics and composites (Print)
(literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#autori
- M.F. Pernice1, D. Annicchiarico2, A. Petriccione2,
G. De Tommaso1, M. Meo1,3, V. Antonucci2 and M. Zarrelli2 (literal)
- Pagina inizio
- Pagina fine
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#url
- http://jrp.sagepub.com/content/30/4/287 (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroVolume
- Rivista
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#pagineTotali
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroFascicolo
- Note
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#affiliazioni
- 1IMAST s.c.a.r.l. - Technology Cluster, Italy.
2CNR, IMCB - Institute for Composite and Biomedical Materials, Italy.
3Material Research Centre, Department of Mechanical Engineering,
University of Bath, UK. (literal)
- Titolo
- Structural polymer composite materials for muscle-skeletal diagnostic systems (literal)
- Abstract
- The aim of the study presented herein was to investigate the possibility of employing composites in the production of
medical diagnostic equipments, and their advantages in terms of costs and performances. In particular, the patientpositioning
system of a magnetic resonance apparatus, comprising two elements, the bed tables, and the footrest, was
examined with the main goal of enabling a wider range of population to be tested, thanks to a full re-designing of their
item. An extensive variety of composites were investigated, in order to select a combination of matrix and reinforcement
able to satisfy the specific requirements in terms of magnetic transparency, structural behavior, manufacturing features,
and costs. Glass fiber composites were selected and characterized by experimental tests. The patient-positioning system
was completely re-designed and analysed by finite elements method; design optimization led to a simplified structure,
offering a series of advantages such as light weight, manoeuvrability, thickness reduction, and cost-saving processes. Final
full-scale prototypes were realized and tested according to the Medical Security procedures, achieving as main results not
only a reduction of the total production cost, but also an increase in the percentage of people that can be scanned by this