http://www.cnr.it/ontology/cnr/individuo/prodotto/ID23253
Control and monitoring system for fusion neutron spectroscopy on the Joint European Torus (Articolo in rivista)
- Type
- Label
- Control and monitoring system for fusion neutron spectroscopy on the Joint European Torus (Articolo in rivista) (literal)
- Anno
- 2004-01-01T00:00:00+01:00 (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#doi
- 10.1063/1.1787934 (literal)
- Alternative label
Tardocchi M., Gorini G., Palma D., Sozzi C., Kallne J., Conroy S., Ericsson G., Giacomelli L., Glasser W., Henriksson H., Hjalmarsson A., Ronchi E., Sjostrand H., Weiszflog M., Popovichev S. (2004)
Control and monitoring system for fusion neutron spectroscopy on the Joint European Torus
in Review of scientific instruments
(literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#autori
- Tardocchi M., Gorini G., Palma D., Sozzi C., Kallne J., Conroy S., Ericsson G., Giacomelli L., Glasser W., Henriksson H., Hjalmarsson A., Ronchi E., Sjostrand H., Weiszflog M., Popovichev S. (literal)
- Pagina inizio
- Pagina fine
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroVolume
- Rivista
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#numeroFascicolo
- Note
- ISI Web of Science (WOS) (literal)
- Http://www.cnr.it/ontology/cnr/pubblicazioni.owl#affiliazioni
- [ 1 ] Univ Milano Bicocca, INFM, I-20126 Milan, Italy
[ 2 ] Univ Milano Bicocca, Dipartimento Fis G Occhialini, I-20126 Milan, Italy
[ 3 ] Uppsala Univ, Dept Neutron Res, EURATOM, VR Assoc, Uppsala, Sweden (literal)
- Titolo
- Control and monitoring system for fusion neutron spectroscopy on the Joint European Torus (literal)
- Abstract
- A new control and monitoring (C&M) system is being developed for the TOFOR and MPRu fusion
neutron spectrometers within the Joint European Torus enhancement program. The system, which is
an evolution of the existing C&M system of the MPR spectrometer, consists of a controlled pulsed
light source distributed by an optical fiber network to all photomultiplier tubes used in the plastic
scintillator based spectrometers. The light source is a green Nd:LSB solid-state laser complemented
by blue light emitting diode sources. Pulse height distributions for each detection channel are
recorded to set the spectrometers to prescribed working points and monitor deviations. Absolute
reference is obtained complementing the controlled light source with radioactive sources. In this
article we report on the C&M prototype design and component tests for the MPRu spectrometer.
The results show that the laser and the associated optics provide a controlled light pulse of intensity
covering a dynamic range of more than four orders of magnitude. The choice of optical fiber
diameters is critical for achieving the desired stability and uniformity of the light intensity collected
by each MPRu detector. (literal)